Comparative Evaluation of Root Canal Working Length Determination with Three Methods: Conventional Radiography, Digital Radiography and Raypex6 Apex Locator: An Experimental Study

Document Type : Original Article

Authors

1 Assistant Professor of Endodontics, Dental Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran.

2 Resident of Endodontics, Endodontics Department, School of Dentistry, Shahid Sadoughi University of Medical Sciences, Yazd, Iran

3 General Denstist, Mashhad, Iran

4 General Denstist, Isfahan, Iran

5 Medical Student, Isfahan University of Medical Sciences, Isfahan, Iran

6 General Dentist, Isfahan, Iran

Abstract

Introduction: Determining working length had always been one of the most crucial factors in evaluating prognosis. Radiography as a gold standard way nowadays has some flaws like making a 3D object, image distortion, not measuring the exact location of apical foramen, and putting the patient in a direct X-ray exposure. Here, we compare these three ways in measuring working length of single canal teeth that are narrow. Methods: Initially thirty single canal teeth with narrow canals were selected. After preparing the access cavity, the teeth were mounted in alginate for measuring working length with an apex locator. After that, they mounted in chalk in order to determine the working length using conventional and digital radiographs. Finally, the teeth were removed from the mount and the exact working length assessed using a hand file to compare with the three mentioned methods. Results: This study showed that the mean measured working length of root canal therapy had a significant difference between the four methods (P=0.003). Bonferroni post hoc test showed that the mean exact working length of root canal therapy was significantly lower than measured working length of root canal therapy by conventional radiography (P=0.002), digital radiography (P=0.001) and Raypex6 apex locator (P=0.01). However, there was no significant difference between these three methods (P>0.05).
Conclusion: The results of this study showed that the mean measured working length of root canal therapy had no significant difference between digital radiography, conventional radiography, and Raypex6 apex locator but these three methods had a significant difference with the exact teeth length

Keywords

Main Subjects


Introduction

Determining working length(WL) had always been one of the most crucial factors in evaluating prognosis(1). One of the most pressing concerns of dentists is determining where is the endpoint of preparation and obturation in root canal therapy (RCT) (2). The working length is the distance between the coronal reference point and the endpoint of the root canal that obturation and preparation ends(3). According to the studies, apical constriction considered the endpoint of the root canal and the most desirable termination point for preparation and obturation. Overextension from the apical constriction demonstrates a poor prognosis(4).

Prevalence of failure at canal working length determination is to the extent that filling the root canal more than 2mm from the radiographic apex or beyond the radiographic apex alongside file fracture and perforation is considered one of the most technical dental failures in root canal treatment. According to the American dental association, the importance of precise apical constriction determination is to the extent that filling the root canal beyond the radiographic apex indicates a technical failure, which can be due to the inaccuracy in working length determination(5).

Preparing the root canal beyond the radiographic apex can transmit microorganisms beyond the apical constriction, which threatens the process of healing periapical tissue. Although in the studies on overfilling with gutta-percha under an electron microscope have shown the possibility of bacterial biofilm(6). This biofilm prevents the immune system from overcoming the bacteria and facilitates their growth, and can cause a foreign body reaction(7).

There are several ways to determine the working length, including apex locators, conventional radiography, and digital radiography, each with its advantages and disadvantages(8, 9). Apex locators widely use besides radiography(10, 11). The best way to determine the working length is radiography(12-14).

Drawbacks of radiography include 2-dimensional shadows on a single film, elongation, shortening, image distortion, and patient exposure(15). Albeit, digital radiography has reduced many of the problems with benefits such as reduced radiation, chemical process removal, speed of obtaining the image, and computer storage(16).

Nevertheless, both conventional and digital graphs still have their disadvantages. Radiographs are unable to delineate the precise location of apical constriction, apical foramen, and CDJ(17, 18).

Apex locators are also contraindicated in patients using Pacemaker(19). It has long been accepted that apex locators should not be used alone and can only be used alongside radiography, which can improve the accuracy of working length determination (20, 21). Apex Locator also reduces the number of diagnostic radiographs to determine the working length, which naturally reduces the radiation dose to the patient(22).

However, in patients whose radiation dose might be harmful to obtain radiographs, an apex locator alone is also recommended for root canal therapy(23). Studies have shown that the apical foramen does not always coincide with the radiographic apex and may be in a lateral position relative to it with a distance of 2 mm(24). Other studies have explicated that the apical constriction can be up to 1 mm away from the apical foramen(25). As a result, radiography is unable to locate the apical foramen in these cases(10).

Apex locator introduced by Suzuki in the 1940s(26). And in 1962, Sunada (27) proposed the idea of ​​using the determination of electrical resistance between the oral mucous membrane and periodontal membrane to define the working length. First and second-generation apex locators in the presence of electrolytes in the root canal cannot accurately determine the working length(28). But third-generation apex locators with more powerful processors show more accurate performance despite electrolytes(29). This study aims to compare the accuracy of working length determination between conventional radiography, digital radiography, and the apex locator (Raypex6).

Methods and Materials

Thirty single-rooted teeth selected based on inclusion and exclusion criteria. The inclusion criteria were as follows: 

  1. Single canal in single-rooted teeth 
  2. Teeth with no evidence of root caries 
  3. Teeth with no excessive curve of the root
  4. teeth with no root fracture
  5. The teeth with narrow root canal 
  6. Teeth with the completely formed apex

Exclusion criteria were as follows: 

  1. Teeth with root resorption
  2. Teeth with root fracture before or during the operation
  3. File fracture during the operation 
  4. Patency with file number #20 or more.

The extracted teeth vertically mounted in chalk and marked from number 1 to 30. The initial file inserted into canals based on tactile sense. Conventional and digital radiography taken with the parallel technique, if the file tip was more than 2 mm away from the radiographic apex, radiography repeated. Each conventional x-ray image evaluated by an endodontist and a dental student by modifying the working length.

If there were more than 1 mm difference between the declared numbers for a tooth, Another endodontist estimated the working length. All three specialists calibrated to define the working length, and the same gauge used during the study. If the lengths reported by the student and the endodontist were less than 1 mm, the mean of the two numbers was recorded as the working length of the tooth according to conventional radiography. This process with Digital radiography repeated for all teeth, except that digital software was used instead of the gage.

For the in-vitro study on apex locator, the teeth embedded into alginate and the lip clip electrode of Raypex6 inserted into alginate neat the tooth. Then the working length estimated by the apex locator(30).

After determining the working length with all three methods, the teeth were extracted from alginate. By inserting file #15 into the canal, tip to tip with the apical foramen, the length of the file measured, and 1mm subtracted from the file length, and the actual tooth length obtained(25).

Lastly, we compared each of the three numbers obtained by conventional radiography, digital radiography, and Raypex6 apex locator for each tooth with the actual length of the teeth. All of these procedures are shown in Figures 1-3.

 

 

 

 

 

 

Figure 1: Method of determining the exact length of the teeth.

 

 

 

Figure 2: Determination of tooth length by conventional radiography

 

 

 

Figure 3: Image of digital radiographs on the screen and determine the length of operation on the computer

 

 

.

 

Data were analyzed with Bonferroni post hoc test using SPSS version 22.

Results

The sample consisted of 30 single-canal teeth with narrow canals, and the working length measured by Conventional Radiography, Digital Radiography, and Apex Locator Raypex6 in millimeters. The outcomes explicated the mean measured the working length of the root canal had no significant difference between the digital radiography, conventional radiography, and Raypex6 apex locator. These three methods had a significant difference with the actual teeth length. (Table I)

 

 

Table I: Mean working length in millimeter 

Method

Mean

Standard Deviation

P-value

Actual length

19.59

1.93

0.003

Conventional Radiography

20.18

1.91

Digital Radiography

20.19

1.92

Raypex6

20.11

2.02

 

 

ANOVA test with repeated observations revealed that the mean working length of the root canal was significantly different between the four methods (P=0.003). Bonferroni post hoc test showed that the mean exact working length of root canal therapy was significantly lower than the measured working length of root canal therapy by conventional radiography (P=0.002), digital radiography (P=0.001) , and Raypex6 apex locator (P=0.01). However, there was no significant difference between these three methods (P>0.05). (Table II)

 

Table II: Comparison of the mean working length of the root canal between two methods by Bonferroni post hoc test

Methods

P-value

actual length and conventional radiography

0.002

actual length and digital radiography

0.001

actual length and Raypex6

0.01

digital radiography and conventional radiography

0.97

conventional radiography and Raypex6

0.68

digital radiography and Raypex6

0.64


 

Discussion

Attaining success in Endodontic treatment depends on multiple factors, one of the most crucial factors is proper root canal preparation, also, keeping the periapical area as intact as possible by determining the exact working length. The ultimate goal of root canal treatment is to fill three-dimensional coronal, apical, and lateral of the canal.

Incorrect measurements can cause overfilling, rupture of periapical tissues, and excessive pain after treatment(31). There are many ways to determine the working length, the most well-known methods are tactile sense, radiography, the use of mean root lengths statistically and, apex locators. However, radiographic interpretation using the initial file is a common and accepted method of estimating the working length(32). The main problem of this method is the variety of anatomy in the apical area.  Also, if apical constriction is not matching the radiographic apex, this method can cause over instrumentation and invasion of the periapical tissues(33). Furthermore, radiographic problems such as image distortion, anatomical structure interference, gag reflex, patient exposure -especially in pregnancy- should consider(34).

One of the most critical issues with root canal treatment is the limited amount of canal space, so the endpoint of operation should define precisely, so that not to damage the apical tissue while cleaning and shaping the root canal. Therefore determining the exact apical constriction is a significant clinical object(32). In various investigations, many scientists determined working length using Electronic apex locators. The results of studies showed that in most cases, there was no significant difference between apex locators. Electronic apex locators also can locate the possible perforation of root or pulp chamber floor, additional canals, position of calcified barrier in apexification(35).

The results of this study explicated the mean measured working length of the root canal had no significant difference between the digital radiography, conventional radiography, and Raypex6 apex locator. Nevertheless, these three methods had a significant difference with the actual teeth length. This significant difference indicates that none of the three methods used can estimate the exact working length.

Elayouti et al. (35) estimated the working length in 182 canals using apex locator Raypex4. The results showed that Apex Locator Raypex4 was reasonably capable of determining the root canal length.

In another study by Wrbas et al., 2007,(36) the accuracy of two apex locator models, Raypex5 and root ZX was compared. Twenty single-rooted hopeless teeth used to determine the working length, and then the initial file fixed with composite resin and teeth extracted. The results explained that Apex Locator devices were able to accurately determine the working length. There was no significant difference between Raypex5 and root ZX.

Vieyra et al. (37). in a 2011 study, evaluated 245 teeth for precision measurement of apex locator and radiologic methods during root canal therapy. The results of the study showed that apex locator devices are more accurate than radiographic methods in determining the working length and can reduce the risk of using tools in dentistry. A study by Orosco et al (38), aimed to compare the accuracy of working length determination, with conventional and digital radiography. The results showed that the accuracy of conventional radiography was higher than digital radiography.

A further study by Kumar et al. (2016) aimed to compare the accuracy of the apex locator and conventional radiography in determining the working length of 41 primary hopeless teeth. The results also showed that there was no significant difference between the apex locator and the radiographic method(39). The differences in various studies are due to the diverse sample size and teeth evaluated in these studies and, the differences with this study are acceptable.

Conclusion

The outcomes of this study explicated the mean measured the working length of the root canal had no significant difference between the digital radiography, conventional radiography, and Raypex6 apex locator. Notwithstanding, these three methods had a significant difference with the actual teeth length. This significant difference indicates that none of the three approaches used can measure the exact working length.

Conflict of interest: none.

Acknowledgment

We would like to express our special thanks of gratitude to the Research Committee of Isfahan University of Medical Sciences for their cooperation and support.

                Goldberg F, Marroquin BB, Frajlich S, Dreyer C. In vitro evaluation of the ability of three apex locators to determine the working length during retreatment. J Endod. 2005;31(9):676-678.
2.             Grove CJ. “Why Root Canals Should Be Filled to the Dentinocemental Junction”**A reply to Dr. Skillen’s criticism published in the November issue.
J Am Dent Assoc (1922). 1931;18(2):314-319.
3.             Dutta K, Desai PD, Das UK, Sarkar S. Comparative evaluation of three methods to measure working length - Manual tactile sensation, digital radiograph, and multidetector computed tomography: An in vitro study. Journal of conservative dentistry : J Conserv Dent. 2017;20(2):76-80.
4.             CHAPMAN CE. A Microscopic Study of the Apical Region of Human Anterior Teeth. Int Endod J. 1969;3(4):52-58.
5.             Lin LM, Rosenberg PA, Lin J. Do procedural errors cause endodontic treatment failure? J Am Dent Assoc. 2005;136(2):187-93; quiz 231.
6.             Gutierrez JH, Brizuela C, Villota E. Human teeth with periapical pathosis after overinstrumentation and overfilling of the root canals: a scanning electron microscopic study. Int Endod J. 1999;32(1):40-48.
7.             Bergenholtz G, Lekholm U, Milthon R, Engstrom B. Influence of apical overinstrumentation and overfilling on re-treated root canals. J Endod. 1979;5(10):310-314.
8.             Seidberg BH, Alibrandi BV, Fine H, Logue B. Clinical investigation of measuring working lengths of root canals with an electronic device and with digital-tactile sense. J Am Dent Assoc. 1975;90(2):379-387.
9.             Stein TJ, Corcoran JF. Radiographic “working length” revisited. Oral Surg Oral Med Oral Pathol. 1992;74(6):796-800.
10.          Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod. 1998;24(6):438-441.
11.          Williams CB, Joyce AP, Roberts S. A comparison between in vivo radiographic working length determination and measurement after extraction. J Endod. 2006;32(7):624-627.
12.          Friedlander LT, Love RM, Chandler NP. A comparison of phosphor-plate digital images with conventional radiographs for the perceived clarity of fine endodontic files and periapical lesions. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2002;93(3):321-327.
13.          Grayli E, Peyvandi A, Mallahi M, Naeemi S, Afshari E. Radiographic Evaluation of Iatrogenic Errors of Root Canal Treatments Performed in an Undergraduate Dental Clinic. J Dent Mater Tech. 2021;10(1):16-21.
14.          Hoseini Zarch SH, Heravi F, Javadian Langaroodi A, Pirgazi H. Evaluation of Cone Beam Computed Tomography in Diagnosis and Treatment Plan of Impacted Maxillary Canines. J Dent Mater Tech. 2013;2(3):92-98.
15.          Goldman M, Pearson AH, Darzenta N. Endodontic success—who's reading the radiograph? Oral Surg Oral Med Oral Pathol. 1972;33(3):432-437.
16.          White SC, Pharoah MJ. White and Pharoah's Oral Radiology E-Book: Principles and Interpretation: Elsevier Inc.; 2018.
17.          Green D. A stereomicroscopic study of the root apices of 400 maxillary and mandibular anterior teeth. Oral Surg Oral Med Oral Pathol. 1956;9(11):1224-1232.
18.          DUMMER PM, McGINN JH, REES DG. The position and topography of the apical canal constriction and apical foramen. Int Endod J. 1984;17(4):192-198.
19.          Pishipati KVC. An in vitro comparison of Propex II apex locator to standard radiographic method. Iran Endod J. 2013;8(3):114-117.
20.          Hembrough JH, Weine FS, Pisano JV, Eskoz N. Accuracy of an electronic apex locator: a clinical evaluation in maxillary molars. J Endod. 1993;19(5):242-246.
21.          Fouad AF, Reid LC. Effect of using electronic apex locators on selected endodontic treatment parameters. J Endod. 2000;26(6):364-367.
22.          Brunton PA, Abdeen D, Macfarlane TV. The effect of an apex locator on exposure to radiation during endodontic therapy. J Endod. 2002;28(7):524-526.
23.          Smadi L. Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: a clinical study. BMC Oral Health. 2006;6(1):1-6.
24.          LAMBJERG-HANSEN H. Vital and mortal pulpectomy on permanent human teeth. Eur J Oral Sci. 1974;82(4):243-332.
25.          Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod. 1993;19(11):545-548.
26.          Suzuki K. Experimental study on iontophoresis. Jpn J Stomatology. 1942;16:411-429.
27.          Sunada I. New method for measuring the length of the root canal. J Dent Res. 1962;41(2):375-387.
28.          Fouad AF, Krell KV, McKendry DJ, Koorbusch GF, Olson RA. A clinical evaluation of five electronic root canal length measuring instruments. J Endod. 1990;16(9):446-449.
29.          Fouad AF, Rivera EM, Krell KV. Accuracy of the endex with variations in canal irrigants and formane size. J Endod. 1993;19(2):63-67.
30.          Kqiku L, Städtler P. Radiographic versus electronic root canal working length determination. Indian J Dent Res. 2011;22(6):777-780.
31.          Kim E, Lee S-J. Electronic apex locator. Dent Clin North Am. 2004;48(1):35-54.
32.          Pratten DH, McDonald N. Comparison of radiographic and electronic working lengths. J Endod. 1996;22(4):173-176.
33.          Cohen S, Hargreaves KM. Pathways of the Pulp. 9 [sup] th ed. St Louis: Mosby. 2006:786-821.
34.          Jafarzadeh H, Beyrami M, Forghani M. Evaluation of conventional radiography and an electronic apex locator in determining the working length in C-shapedcanals. Iran Endod J. 2017;12(1):60-63.
35.          ElAyouti A, Kimionis I, Chu AL, Löst C. Determining the apical terminus of root‐end resected teeth using three modern apex locators: a comparative ex vivo study. Int Endod J. 2005;38(11):827-833.
36.          Wrbas K, Ziegler A, Altenburger M, Schirrmeister J. In vivo comparison of working length determination with two electronic apex locators. Int Endod J. 2007;40(2):133-138.
37.          Vieyra J, Acosta J. Comparison of working length determination with radiographs and four electronic apex locators. Int Endod J. 2011;44(6):510-518.
38.          Orosco FA, Bernardineli N, Garcia RB, Bramante CM, Duarte MAH, Moraes IGd. In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX.
J Appl Oral Sci. 2012;20(5):522-525.
39.          Kumar LV, Sreelakshmi N, Reddy ER, Manjula M, Rani ST, Rajesh A. Clinical evaluation of conventional radiography, radiovisiography, and an electronic apex locator in determining the working length in primary teeth. Pediatr Dent. 2016;38(1):37-41.